Common Practices and Procedures Not Found in
the Text
Experiments are reported an informal style, yet
the information is complete. Organization of the data provides
not only a clear but also a logical report. In the Example Tables
below a format is given that is used in several of the
experiments in this and future courses.
Example Reaction Table
This is a reaction with stoichiometry more
complex than you will encounter in this course, but the example
is instructive. The ultimate goal is to determine the theoretical
yield, given the amounts of reactants. Pyridine (C5H5N)
is converted with CrO3 and HCl to form an intermediate
compound, pyridinium chlorochromate, that is then used to oxidize
1-decanol (C9H19CH2 OH) to
decanal (C9H19CHO). The overall reaction
may be written without needing to write the formula for
pyridinium chlorochromate: C5H5N + CrO3
+ HCl + C9H19CH2 OH --> C9H19CHO
+ CrCl3 + H2O + C5H5NHCl
and when balanced: 2C5H5N + 2CrO3
+ 8HCl + 3C9H19CH2 OH -->3C9H19CHO
+ 2CrCl3 + 6H2O + 2C5H5NHCl.
In this example the procedure calls for 1.0 mL of pyridine, 4g of
chromium trioxide, 6 mL of 6M HCl and 1.0 mL of decanol.
| 2 C5H5N | 2 CrO3 | 8 HCl | 3 C9H19CH2OH | -> 3 C9H19CHO | |
| vol. used | 1.0 mL | 6 mL (aq) | 1.0 mL | ||
| density | 0.978 g/mL | 0.829 g/mL | |||
| mass | 1.0 mL x 0.978 g/mL =0.978 g = 978mg |
4g = 4000mg | 1.0 mL x 0.829 g/mL = 0.829g = 829 mg |
5.2 mmole x156 mg/mmole = 811mg = theoret. yield |
|
| MW | 79 mg/mmole | 100 mg/mmole | 158 mg/mmole | 156 mg/mmole | |
| Molarity | 6 Molar = 6 mmole/mL |
||||
| mmoles | 978 mg ÷79 mg/mmole =12.0 mmole |
4000 mg ÷100 mg/mmole = 4.0 mmole |
6 mL x6 mmole/mL = 36 mmole |
829 mg ÷158 mg/mmole = 5.2 mmole |
5.2 mmole |
| calc. lim. reagent |
12 ÷ 2 = 6 | 4.0 ÷ 2 = 2 | 36 ÷ 8 = 4.5 | 5.2 ÷3 = 1.7 limiting reagent |
Densities, molecular weights are obtained from
the Handbook. Note that many of the blanks are not filled in
because the data is not necessary. Also the mmoles of HCl are
calculated from a VxM = moles calculation since HCl is in aqueous
solution. The "calculate lim. reagent" line is usually
not necessary, but since the ratio of coefficients in the
balanced equation is not 1:1:1 etc., another line calculation
must be made; when the millimoles of decanol is divided by its
coefficient the result is lower than the result for any other
reactant, therefore decanol is the limiting reagent.
Note also how the calculations are made
right in the table. The logic is followed with arrows so that the
limiting reagent and the theoretical yield, both labeled, are
easily found.
Example Table of Chemical Tests.
This table is appropriate for the
identification of unknowns. Observe how the IR information is
included into the line of reasoning and suggests the chemical
tests. Even though the IR clearly indicates an alcohol,
preliminary alkene and alkyl halide tests are run anyway, since
negative tests provide confirmation. But not all chemical tests
need be run, only the ones suggested by the IR and preliminary
tests. The alkyl halide (Beilstein) test, like the Lucas test,
was not decisive the first time, and tests were run to check the
initial findings. The road to determining the functional group is
often laborious. Expect conflicts and to repeat tests.
| Test | Result | Inference | Comments |
| IR 3200-3600 cm-1 | rounded peak - large - bottoms @ 3350 cm-1 |
alcohol | |
| IR 3100-3000 cm-1 | no peak -only a shoulder | no sp2 C-H | not an alkene - not aromatic |
| IR 3000-2800 cm-1 | peaks | sp3 C-H | alkyl |
| IR 1680-1630 cm-1 | nothing more than noise | probably not C=C | confirms 3100-3000 cm-1 test |
| IR 1000-1200 cm-1 | rounded peak 1060 cm-1 | primary alc. | see Mayo, Table 6.12 |
| 1H NMR number of signals | at least 5 | at least 5 uniquely positioned H's in the molecule | |
| 1H NMR 5.0-6.5 ppm | none | no H-C=C | confirms above - not an alkene |
| (other 1H NMR tests) | |||
| 13C NMR number of "all carbon" signals | at least 4 | at least 4 uniquely positioned C atoms in the molecule | |
| 13C NMR number of 100-150 ppm signals | none | no C=C | confirms above - not an alkene |
| (other 13C NMR tests including DEPT) | |||
| Boiling Point | 151-155 deg Celsius | ||
| Beilstein | green flame | alkyl halide | alcohol which also has halogen ? |
| Br2/CCl4 | red persists | not an alkene | confirms IR |
| BP repeat | 158-160 deg Celsius | if alcohol - about 5-6 carbons | first time was less clear - this time I took my time |
| Lucas | no cloudiness -did not dissolve | more than 6 carbons | |
| (Beilstein on known pentanol) | green flame but very brief | false positive | recheck unknown |
| Beilstein repeat on unknown | very brief green | probably not alkyl halide | green not persistent |
| (Beilstein on known chlorobenzene) |
much stronger and longer green | real positive | unknown is definitely not like chlorobenzene that really has halogen |
| Chromic Acid | opaque blue-green suspension | primary or secondary alcohol | |
| ceric nitrate | yellow to red | alcohol of less than 10 carbons | From Lucas, BP and this, unknown is an alcohol having 5-10 carbon atoms |
| Redo Lucas using fresh reagent |
cloudiness in 3 min | secondary alcohol | other students noticed that the Lucas was not working - made a fresh solution. Therefore unknown can be 6 or less carbons? |
| Repeat Lucas using fresh reagent | cloudiness in 2.5 min | secondary alcohol | IR indicates primary. I cleaned the test tubes and pipettes. |
| Iodoform | (didn't have time) | not that important - only certain secondary alcohols will be positive |
Example Table of Candidates.
This table is made to organize the selection of
a single compound from a list that best fits the data. In this
example the progress is followed from a Table of Tests that
determined that the unknown was an alcohol having about 6 carbons
and was either primary (from the IR - but note that the 1060 cm-1
is a little high) or secondary (from the Lucas). So all
candidates alcohols having boiling points 5 degrees below and 15
degrees above 159 are listed.
| Candidate | Synonym | Structure | 1o 2o 3o | bp (oC) | Why Does This Not Fit My Spectrum ? |
| (My unknown) | ~164 | ||||
| n-hexyl alcohol | 1-hexanol | CH3CH2CH2CH2CH2CH2OH | 1o | 157.5 | peak @ 1050 cm-1too wide |
| 2-hepanol | CH3CH2CH2CH2CH2CH(OH)CH3 | 2o | 158 | has 1100 cm-1 peak -no | |
| cyclohexanol | [you must supply this drawing] | 2o | 161.5 | best fit - small extra peak at 930 cm-1 | |
| 2-methyl-1-hexanol | CH3CH2CH2CH2CH(CH3 )CH2OH | 1o | 164 | has large 1130 cm-1 - no | |
| 2-methylcyclohexanol | [you must supply this drawing] | 2o | 167.4 | 1350 cm-1peak
too small. Too many small peaks bet. 1050-1100 cm-1 |
|
| n-heptyl alcohol | 1-heptanol | CH3CH2CH2CH2CH2CH2CH2OH | 1o | 176.8 | 1380 cm-1 peak is too small. 1050 cm-1 is too fat |
All candidates must have simple drawings of
their structures (only some, above, could be printed in this
example.) Structures are found in the Handbook in an appendix
"Structural Formulas of Organic Compounds" to the list
"Physical Constants of Organic Compounds"
A number of techniques are introduced in the
first laboratory. Below are some more detailed instructions that
will be referred to in later laboratories.
Handling of Liquids. See Mayo, et al., p.40-45 Microscale amounts of liquids are not poured but instead transferred with a pipette. In your drawer are Pasteur pipettes and rubber bulbs - which is used for most work; also in the drawer are a graduated pipette (which fits into the green pi-pump) and a syringe.
As explained in the Check-In lab, Pasteur pipettes are held in the palm of the hand with the smallest finger while the rubber bulb is squeezed with the thumb and forefinger. This technique gives the best control of the pipette tip when the other hand is holding the vessels holding the liquids.
During any transfer with a pipette some liquid may leak out. To prevent loss of this liquid it is important to bring the mouths of the dispensing and receiving vessels as close as possible together.
The Pasteur pipette holds about 2 mL and is easily cleaned for further use. If more than a few milliliters of acetone are not enough to clean the pipette it is cheaper to throw it away and replace it from the stockroom.
The 2 mL graduated pipette/pi-pump combination
measures with more accuracy, but when you hold it with your thumb
on the roller you will see that the pipette tip is unsteady. The other hand may need to hold the tip, but this will
not allow you to hold the source or destination containers. In
addition, the graduated pipettes are not so cheap to replace.
Handling of Solids. Microscale
quantities of solids are weighed on ordinary pieces of paper; but
these can be folded to channel the solids as they are then poured
into the test tube, flask, etc. containers. Because small amounts
are used it is advantageous to precut the paper so the solid runs
along a fold directly to the bottom of the vessel without
touching its neck. For emptying vessels use of the microspatula
requires practice and patience.
Boiling Point. The micro technique will
be described in a future procedure.
Recrystallization. This is a technique
for purification of solids, Generally, a solvent is selected
that, when boiling, will dissolve the solid. When the solution is
then cooled, the solvent will not be able to dissolve the solid
as well. As a consequence, purified solid will precipitate out of
solution as crystals. These crystals are recovered by filtering
them from the solvent; the separated solvent carries the
impurities with it. Specific instructions will be given in
procedures.
Routine use of the Gas Chromatograph. Text p. 66-70. Instrument is a Gow-Mac with a Carbowax 20 M (polar) column and a DC-200 (non - polar) column. Setting up an instrument such as the gas chromatograph is a skill learned in an advanced course, but the parameters of the analysis must be measured and written right on the chart: The carrier gas, helium, with a back pressure at about 18 pounds per square inch read at the gauge and a flow about 20-25 ml/min read at the flowmeter. The column used, the oven temperature, the attenuation, the nature of the sample and the amount injected, the attenuation (i.e. the sensitivity of the detector) and the chart speed. (about 1 inch per minute; this should be verified with a ruler and a clock.). The instrument will be set up for your work; make sure that the detector power is on only when helium is flowing.
Prior to your injection, adjust the pen, with the chromatograph zero knob, to the right side of the main chart (the small chart on the right side of the recorder is not used.)
To inject, take the automatic syringe and clean it several times by filling and emptying it several times with your sample; discharge the syringe into the air. Adjust the barrel of the syringe to the microliter (µL = 0.001 L) capacity desired. For normal runs 3 µL is typical. Take up the sample into the syringe, and inject it into the injector port corresponding to the desired column. Make a mark with a pen on the chart at the point of injection. While you await peaks, make sure you have documented the parameters underlined above on the chart.
The first injection may cause the chromatogram to be too big or too small. To make the peaks larger, increase the sensitivity by lowering the attenuation. If your peaks of interest are too large, turn up the attenuator. Often large peaks which are not of interest, are allowed to pin the pen so that smaller peaks can be seen and measured. This will be the case with the free radical chlorination experiment.
If another student is about to inject, permit him/her to inject and wait until your chromatogram has passed the sprocket. Now tear off your chart, take it to your bench and staple it to an empty page in your notebook. Identify as many of the peaks as possible by marking them. The instructions for measuring the retention time and integrating each peak are given in the "check-in" lab.
Calculations of the retention times for the peaks of interest must be put on the chromatogram or on the same notebook page that the chromatogram is attached to. Since the chart speed is 1 inch per minute, it is easiest to measure (in millimeters) the horizontal distance between the injection point and the point when the peak has reached its maximum. Multiply this distance by (minute/25.4 mm) to obtain the retention time. Repeat this for all peaks needing to be measured.
Calculations for the areas for each peak to be measured - also written on the chromatogram - are done by the height times width-at-half-height method: 1) draw a baseline that would have resulted if the peak had not appeared; 2) from the baseline measure and record the height, the vertical distance in millimeters from the baseline to the top of the peak; 3) mark the midpoint of this line; 4) measure the width of the peak, through the midpoint mark (i.e. the width-at-half-height). Multiply the height by the width-at-half-height and record the result. The area of each measured peak is then calculated as a percent of the total measured peaks.
Preparation and Analysis of Samples for an Infrared (IR) Spectrum . McM sections 12.5-12.9. (the Mayo, et al. treatment in Ch. 6 need only be scanned since it is overly detailed for an introduction). The procedure the preparation and measurement of a liquid (neat) sample is found in the "check-in " lab file.
The instrument is a Perkin Elmer Model Spectrum One infrared spectrometer.
For liquid samples disk shaped NaCl plates are used to place the sample in the beam.
| a) | A blank "background" scan is run first to set the instrument. If it has not already been done by your predecessor, stand each plate on its edge on a paper towel and squirt both faces of the salt plates with about 2 mL of 100% ethanol. (Never wash the plates with water - or anything containing water such as soap - as they will dissolve. Salt plates cost $25 each..) The salt plates are mounted on a holder somewhat like Mayo Figure 6.43 except that they are not screwed down but instead held in place by a metal retainer. Run a blank "background" spectrum by placing the sample holder (containing the two clean plates) in the sample beam aperture. |
| b) | Remove the sample holder and return to the laboratory to remove the plates and place 1/2-1 drop (30-50 µL) of your liquid sample on one of them. Cover it with the other plate, and remount the plate, and return the holder to its mounting in the instrument. |
| c) | Begin "scan" after filling out the information on the screen. The "Name" of the sample is unknown (fill in "Unknown # ...") The "Sampling Method" for this procedure is "neat, between NaCl plates". The "Analyst" is (fill in your name). The "Fraction" entry is left blank this time. The date and time of the scan will automatically be noted in the printed spectrum |
| d) | Print one copy of the spectrum. Remove the holder then the plates as before and clean the plates with ethanol (remember - no water or soap) for the next student who starts by running the blank. |
Prepare the spectrum for interpretation by inserting and identifying a vertical line in your spectrum that marks 3000 cm-1. This line will be helpful in identifying your unknown since any sharp peaks to the left of this line (between 3020-3100 cm-1 - see McM Table 12.1) represent C-H bonds in which the C is sp2 hybridized. Absorptions to the right of this line, 2960-2850 cm-1 represent C-H bonds in which the C is sp3 hybridized.
Solids melting below 80 oC are run as "melts": a clean salt plate is loaded with 3-5 mg of the solid and the plate is carefully warmed on a clean, dry surface (aluminum foil on a warm hot plate) until the solid is melted. The sample is mounted and run as usual.
Solids melting above 80 oC are run as KBr pellets: 1) obtain a small agate mortar and pestle from the stockroom, and grind 3-6 mg of sample and about 50 mg of KBr until the solid is caked and glassy. 2) With a spatula, transfer the mixture to a cup provided for reflectance sampling. The instructor will then press the sample into a window that will be mounted in the spectrophotometer and the sample is run as usual.
Interpreting an IR Spectrum. In the
"check-in" lab, identification of the unknown substance
only requires comparison of the spectrum obtained with 25 spectra
in chapter 6 of Mayo, et al.. All future spectra must be
interpreted and annotated even if a match happens to be found. In
McM, section 12.5 you will find a quick introduction to IR
theory: vibrations of the bonds within molecules give rise to the
various peaks in IR spectra, and therefore analysis of the
spectra will give information on the bond sequences or functional
groups within the compound being scanned.. Mayo, et al., p. 231,
gives a good strategy on organizing this information.
1. Measure and mark the wavenumber values
(corrected, if necessary, after calibration) for all significant
peaks between 4000 and 1350 cm-1. Remember that the
wavenumber values, not so much the intensities, of these peaks is
important. Also note the shapes (rounded, sharp, wide, narrow) of
the peaks.
2. If solvents or Nujol were used to mount the sample, they must be marked out. Spectra for these substances are found in the Aldrich IR Library
3. Interpretation of the spectrum itself is best done by first checking major features for fundamental functional groups. There are 3 regions to scan initially. These are best followed from McM Figure 12.14 and the figures in Mayo, et al., Chapter 6:
| a) | the O-H and N-H stretching region between 3600 and 3200 cm-1 An alcohol is very easily detected by a very large parabolic peak in this region (Fig 6.24), while an amine N-H peak is less rounded and dominating and for primary amines appears as two peaks (Fig 6.32 - see also Fig 6.35). The carboxylic acid O-H peak is entirely different; it is rounded and strong, but it is also very wide, extending from 3500 to 2500 cm-1 (Fig 6.29) |
| b) | the C-H stretching region between 3100 and 2850 cm-1 is conveniently divided down the 3000 cm-1 line. Almost invariably the peaks between 3000 and 2850 are due to bonds between H and sp3 hybridized carbon atoms; this H-C stretch is present in every figure in Chapter 6 except Figs 6..10b (DCCl3) and 6.39 (chlorobenzene). Absorptions between 3100 and 3000 cm-1 arise from bonds between H and sp2 hybridized carbon, as in alkenes (Fig 6.15) and aromatic compounds (Fig. 6.39. Chlorobenzene is aromatic because it contains a benzene ring, a group of 6 carbons bound in a ring with alternating double and single bonds. Details of this group will be given in Chem 3020, but its spectral characteristics are easily recognized). |
| c) | the carbonyl (C=O) stretch between 1690 and 1760 cm-1. This feature is in carboxylic acids, aldehydes, ketones, esters and amides. |
| d) | other distinguishing peaks are in McM table 12.1. The functional groups listed are of less importance than a) - c) above. Most of the functional groups listed has an example in one of the Figures in Mayo, et al., Chapter 6. |
4. The Handbook has "Infrared Correlation Charts" which are helpful. In the Aldrich IR Library the spectra are organized by functional group (alkenes, alcohols, etc.); each section begins with a general discussion of the characteristic peaks that are present in the functional group.
5. Remember to look first at the most outstanding aspects of the spectrum, not get into minor details. Do not try to interpret every peak in the spectrum.
6. Also remember that the absence of a peak is often just as important as finding one, for it will narrow the list of possibilities for your compound.
7. Now that the major peaks have been identified, further confirming and sub-classifying peaks can be located. The best strategy for this is to refer to the appropriate section in McM. Always check example spectra in McM and in Mayo, et al. Details of groups that may be encountered in this course:
| a) | alcohols - Mayo Table 6.12. Examination of the C-O stretching peak is useful for narrowing the alcohol to either a primary, secondary or tertiary alcohol (these figures often do not apply to cyclic alcohols). |
| b) | alkenes - McM Section 12.8. |
In general, once a functional group is
identified, specifics on its IR characteristics can be confirmed
and sub-classified by looking up the group's IR data in the
"spectroscopic analysis" section of the appropriate
chapter of McM.
8. A match can be sought from an on-line reference: http://www.aist.go.jp/RIODB/SDBS/menu-e.html
or The Aldrich Library of FT-IR Spectra by C.J.
Pouchert, Aldrich Chemical Publishing Co. Milwaukee, WI. For an
exact match there should be a peak for peak correspondence
between your spectrum and the library entry. Note, however, that
the intensities will differ if the amounts sampled are not
the same. Also the wavenumber values should match even though
scaling is different; the spectrum you obtain is linear in
wavenumbers (cm-1 values) while the Aldrich Library
spectra are linear in microns, so each major peak should be
measured separately. A handy conversion: wavenumbers =
10,000/microns.
9. Mark your spectrum liberally. Make notations
on each diagnostic peak (see McM Figure 12.13 for example) and
for each peak that you used to arrive the structure of your
unknown.
10. If two candidate compounds you found in the
Aldrich Library are both almost identical to your spectrum, make
a peak-for peak comparison. To do this mark off two 0.5-1 cm-wide
bands along the length of your spectrum. Label each band is
labeled with a candidate's name and insert small vertical lines
in the positions corresponding to that candidate's peaks (heavy
lines for strong peaks, etc.) All the information is now on your
spectrum and you should be able to determine the best fit for
your unknown.
References:
Text, frequently "Mayo" in the procedures: Mayo, D.W., Pike, R.M., Butcher, S.S. and Trumper, P.K. Microscale Techniques for the Organic Laboratory; Wiley: New York, 1991
Text for the Lecture, frequently "McM" in the procedures: McMurry, J. Organic Chemistry, 5th ed., Brooks/Cole Publishing Company, Pacific Grove, CA. 1999
The "Handbook" refers to any of the recent editions of: Weast, R.D. Handbook of Chemistry and Physics; The Chemical Rubber Co.: Cleveland, 1960-present.
The "Aldrich IR Library"
refers to any edition of: Pouchert, C.J. The Aldrich Library
of Infrared Spectra; Aldrich Chemical Co. Milwaukee,
1970-present.
Rev. January, 2001