I. Digital Micropipettors
A digital micropipettor is a precision pump fitted with a
disposable tip. The volume of air space in the barrel is adjusted by screwing the
plunger in or out of the piston, and the volume is displayed on a digital
readout on the top upper portion of the pipette. Depressing the plunger
displaces the specified volume of air from the piston; releasing the plunger
creates a vacuum, which draws an equal volume of fluid into the tip. Depressing
the plunger again then expels the withdrawn fluid.
There are different size micropipettors with various volume range: a small-volume (1-20 ul), a mid-range (20-100 ul), and a large-volume (100-1000 ul). Depending on the volume needed, appropriate size micropipette should be used to ascertain accuracy. Since these micropipettors are very delicate, hard to repair and very costly, please note the following rules in using a digital micropipettor.
• Never rotate the volume adjuster beyond the upper or lower range of the pipette as stated by the manufacturer.
• Never use the micropipettors without the tip in place; this could ruin the piston.
• Never invert or lay the micopipettor down with a filled tip; fluid could run back into the piston.
• Never let the plunger snap back after withdrawing or expelling fluid. This could damage the piston.
• Never immerse the barrel of the micropipettor in fluid.
• Never reuse a tip that has been used to measure a different
reagent.
General use of digital micropipettors:
1. Rotate the volume adjustor to the desired setting. Be sure
to locate the decimal point when reading the volume setting.
2. Firmly seat a proper-sized tip on the end of the
micropipettor.
3. When withdrawing or expelling fluid, always hold the tube
firmly between your thumb and forefinger. For best control, grasp the
micropipettor in your palm and wrap your fingers around the barrel; work the
plunger with your thumb.
4. The micropipettor plunger has two positions. Depress the
plunger to the first stop and hold it in this position. Dip the tip into the
solution to be pipetted and draw fluid into the tip
by gradually releasing the plunger. Be sure the tip remains in the solution
while you are releasing the plunger.
5. Slide the pipette tip along the inside wall of the reagent
tube to dislodge any excess droplets adhering to the outside of the tip. Check
and make sure there is no air space at the very tip of the tip.
6. To expel the sample into a reaction tube, slowly depress
the plunger to the first stop to expel the sample. Depress to second stop to
blow out the last bit of fluid. Hold the plunger in the depressed position and
slide the tip out of the pipette along the inside wall of the reaction tube.
This creates a capillary effect that helps draw the last bit of fluid out of
the tip.
7. Eject the tip by depressing the tip-ejection button.
8. Use a fresh tip for each new reagent to be pipetted.
This exercise simulates setting up a reaction, using a micorpipettor with a range of 20-200 ul.
1. Use a wax pencil to label three 2-ml tubes A, B, and C.
2. Add
solutions to these tubes using the following table:
|
Tube |
Solution 1 |
Solution 2 |
Solution 3 |
Solution 4 |
|
A |
10 ul |
20 ul |
20 ul |
0 ul |
|
B |
30 ul |
10 ul |
0 ul |
10 ul |
|
C |
10 ul |
0 ul |
20 u1 |
20 u1 |
3. A total of
50 ul of reagents was added to each tube. To check the accuracy of your measurements,
set the pipette to 50ul and withdraw the solution from each tube.
a. Is the tip barely filled?
b. Does a small volume of fluid remain in the
tube? This indicates an over-measurement.
c. After redrawing all fluid, is a space left in the
end of the tip? This indicates an
under-measurement.
d.
If several measurements were inaccurate, repeat the exercise to obtain nearly
perfect results.
Measurement of blood glucose levels was among the
first chemical procedures employed in clinical laboratory medicine. Keilin and Jartree introduced glucose oxidase
methodology in 1948. Keston
later reported use of the combined glucose oxidase-peroxidase
reagent, followed by the Teller addition of chromogenic
reagent to Keston's procedure.
Glucose is oxidized in the presence of glucose oxidase (GO). The
hydrogen peroxide formed reacts, under the influence of peroxidase,
with phenol and 4-aminophenazone to form a red-violet quinone
complex. The intensity of the color is
proportional to glucose concentration.
Glucose + O2 + H2O ------> gluconic acid + H2O2
H2O2 + 4-aminophenazone +
phenol -------> quinone complex
1. Use one standard solution to make diluted solutions
for standard curve.
2. Explain the physiological significance of glucose in
the blood and why abnormal levels (too high or too low) are clinically
significant.
3. To determine how Beer’s law is used to determine the
concentration of molecules in a solution.
4. Use the graphic and formula method to determine the
concentration of glucose in an unknown sample.
Microliter pipette (6) and pipette tips.
Constant water bath set at 37oC.
Colorimeter (6)
Cuvettes (6sets of 7)
Glucose --250 ml @ 400mg/dl
Unknown glucose solution (150 mg/dl-- 200 ml) and glucose reagent.
The standard curve should intersect the origin (zero concentration = zero absorbance).
28. Determine the concentration of the unknown using the standard curve.
29. Locate on the y axis the absorbance of the unknown.
23. Then, use a ruler to draw a horizontal line from this point to its intersection with the data line (standard curve).
24.
Now extend a vertical line to x axis; this will give the concentration of the
unknown.
Beer's law calculation.
Values are derived by the following equation:
Glucose (mg/dl) = Au/As*100,
Where Au = absorbance value of unknown (tube
6)
As = absorbance of standard (tube 7)
and 100 =
concentration of the standard (100 mg/dl)
Normal
range: serum/plasma = 70-105 mg/dl
Reference:
1. Fox,
S.A. Laboratory Guide-
Human Physiology. 7th ed. Pp 28-32,1993.
Wm. C. Brown Publishers, Dubuque, Iowa.
2. Bloom,M.V., Freyer,G.A.
and Micklos, D.A.
1996 Laboratory
DNA Science.
The Benjamin/Cummings Publishing Company, Inc.